Abstract

Thin-film heat flux gauges (HFGs) have been used for decades to measure surface temperatures and heat flux in test turbines with the majority being used in facilities that are short-duration. These gauges are typically composed of two resistive temperature devices deposited on opposing sides of a dielectric. However, because these sensors have been traditionally applied for measurements in transient-type facilities, the challenges facing adaptation of this technology for a steady facility warrant investigation. These challenges are highlighted, and the solutions are presented throughout the paper. This paper describes the nanofabrication process for heat flux gauges and a new calibration method to address the potential deterioration of gauges over long runtimes in continuous-duration facilities. Because the primary uncertainty of these sensors arises from the ambiguity of the thermal properties, the emphasis is placed on the property determination. Also, this paper presents a discussion on the use of impulse response theory to process the data showing the feasibility of the method for steady-duration facilities after an initial settling time. The latter portion of the paper focuses on comparing well-established heat flux gauges developed for short-duration turbine test facilities to recently developed gauges fabricated using modern nanofabrication techniques for a continuous turbine test facility. The gauges were compared using the test case of an impinging jet over a range of Reynolds numbers. The comparison between the PSU gauge and the reference device indicated agreement within 14%, and similar results were achieved through comparison with established sensors from partner institutions.

Introduction

For nearly four decades, thin-film heat flux gauges (HFGs) have been used to quantify heat transfer metrics for turbine components. The initial development of these sensors for turbomachinery was led concurrently by Oxford University [1,2] and MIT [3]. In the years that followed, this technology expanded to other institutions, including The Ohio State University [4], The von Karman Institute (VKI) [5], Virginia Tech [6], and The U.S. Air Force Research Laboratory (AFRL) [7,8]. For each of these users, the HFGs are applied in short-duration, blow-down tests with total operating times on the order of seconds or less. In contrast, continuous-duration facilities that operate with time scales of hours and days pose potential challenges for installation and operation of HFGs as well as post-processing.

This disparity of runtimes is illustrated in Table 1 showing a representative single-test runtime for each facility. Using these runtimes, a worst-case scenario is calculated assuming three tests per day operating 365 days per year over a period of 20 years. For many of these facilities, a cumulative 20-year test time is on the order of a few hours, a period that will easily be overcome in a single-test day for a continuous-duration facility such as the Steady Thermal Aero Research Turbine (START) facility at the Pennsylvania State University. In reality, most of these test articles can only operate one or two tests per day over a maximum of about 4 days per week, therefore further widening the gap.

Table 1

Comparison of HFGs in a range of rotating facilities

FacilityRefLength of test (s)Tests (day)Cumulative 20-year test time (h)
Ohio State TTF[9]0.12030.730
MIT[10]0.20031.217
Oxford ORF0.20031.217
Oxford OTRF0.50033.042
VKI[11]0.50033.042
AFRL2.500315.208
Virginia Tech[6]25.0003152.08
Penn State8 (h)158,400
FacilityRefLength of test (s)Tests (day)Cumulative 20-year test time (h)
Ohio State TTF[9]0.12030.730
MIT[10]0.20031.217
Oxford ORF0.20031.217
Oxford OTRF0.50033.042
VKI[11]0.50033.042
AFRL2.500315.208
Virginia Tech[6]25.0003152.08
Penn State8 (h)158,400

This paper focuses on understanding the challenges associated with implementing HFGs in a continuous-duration facility and presenting viable solutions. A new in situ calibration procedure is evaluated, and the influence of thermal properties on calculated parameters is assessed. For these cases, an impulse response filter is used and considered for use in continuous-duration conditions. Finally, tests conducted with sensors manufactured at the Nanofabrication Lab at Penn State are compared with a commercially available heat flux microsensor (HFM) as well as established gauges currently in use at Oxford and AFRL.

Background

Epstein et al. [3,10] outlined the theory for double-sided heat flux gauges, followed by other practical studies on mid-span turbine heat transfer using the instrumentation with numerical post-processing schemes [12,13]. Since then, this technology has enabled the study of unsteady heat flux for turbine components in both the stationary [5,14,15] and rotating [1517] reference frame of airfoils, in the tip region [1820], and more, benefitting from a minimally intrusive design. Different types have been used [2124], but all operate by solving the unsteady conduction equation through a substrate or substrates given a set of boundary conditions.

These boundary conditions can be obtained in two different ways. For example, a single-sided gauge, as seen in Fig. 1(a) solves the conduction equation using one measured temperature and applying a semi-infinite approximation. In contrast, a double-sided thin film heat flux sensor, Fig. 1(b), has a temperature measuring device on both sides of a known substrate.

Fig. 1
Typical heat flux gauge designs: (a) single-sided and (b) double-sided
Fig. 1
Typical heat flux gauge designs: (a) single-sided and (b) double-sided
Close modal

A double-sided thin film heat flux sensor must be used when operating in steady conditions because the backside boundary condition cannot be treated as semi-infinite. By adding the second sensor, one can always measure the heat flux because the backside boundary condition is measured. This property makes it the right choice for a steady facility. To measure heat flux accurately, there must be a measurable temperature difference across the substrate. Because the substrate is inherently thin for this type of sensor, large heat fluxes are necessary to create a temperature difference discernable from the uncertainty in the measurement. For the START Lab, this large heat flux comes from actively cooling the part which creates an expected gas-to-metal temperature ratio of 1.3 K/K found from CFD predictions.

Although other types exist, the focus of this study is on single- and double-sided thin-film HFGs where the dielectric beneath the temperature sensing element is polyimide. This limited scope is due to the way in which the instruments have been used in the past and will be first used in the START facility. However, as sensor fabrication becomes ever more intricate (such as direct deposition to an airfoil [25]), the solutions in this paper to the challenges in adapting heat flux gauges to a steady measurement will still hold. To obtain the temperatures required to solve the unsteady conduction equation, thin-film resistive temperature devices (RTDs) are often implemented. To create these sensors, the manual application of platinum paint may be used, but lithography processes can significantly improve the control of dimensionality and alignment for double-sided gauges. A lithography technique typically requires advanced equipment but has been used to successfully execute gauge designs using nickel- [3,26] and platinum- [5,27,28] resistive elements.

Beyond sensor fabrication, material properties of the dielectric must be precisely known to achieve accurate calculated results. This study focuses on a polyimide due to its dimensional stability, dielectric properties, and flexibility, but other materials such as quartz [29] and Macor [30] have been used in the past. Most important, it has been previously identified that the bulk thermal properties for polyimide substrates provided by the manufacturer can vary by as much as 20% from the measured value and therefore must be independently validated [3].

In some cases, previous studies have empirically quantified these thermal properties as lumped parameters, specifically ρcpk and k/d [24]. As one example of this method, a known heat flux can be introduced to the HFG and the sensor response can be used to deduce the parameters of interest. Unfortunately, identifying a viable heat flux source can be difficult. Past studies have used an oil bath [26], a laser [3], and even the resistive element itself as the heater [31]. However, many authors have converged on a heated air gun serving as an impinging jet on the gauge for its repeatability and accurate representation of convective experimental conditions [24,26,32].

Fabrication

Within Penn State’s Nanofabrication Lab, a process was developed to produce single- and double-sided HFGs with high yield and repeatable results. The workflow is similar to that presented by Collins et al. [28]. Following this process, the gauges were manufactured through a combination of subtractive and additive processes, starting with a sheet of Pyralux®, a commercially available product comprising a polyimide sheet with 9 µm of copper on each side. In the first part of the fabrication process, copper is removed from the Pyralux through an etching process, creating the pattern for the lead wires. The substrate then goes through a patterning process involving a mask for the platinum elements. Platinum is then deposited through an evaporative physical vapor deposition process, which differentiates this process from Collins et al. [28], where the platinum was sputtered. The platinum bridges the two lead wires and functions as the RTD element.

After completing these steps, a similar process is completed on the opposite side of the substrate. Penn State’s capabilities enable front and back side feature alignment within 5 µm. Through an application of the latest nanofabrication techniques, this alignment capability represents a new standard for HFG fabrication, a factor of two improvements over original manufacturing guidelines reported by Epstein et al. [10].

Accurate alignment capability is important for an orthogonal one-dimensional (1D) heat flux assumption to hold. Because the metal temperature is not a fixed parameter in the START turbine, large spatial gradients of temperature across the part could lead to a breakdown of the 1D assumption for these gauges. By ensuring the gauges are aligned to within 5 µm (5% of the smallest geometrical dimension for the present design), the heat flux can be treated as one-dimensional and orthogonal to the airfoil surface.

Figure 2 compares the design intent to the fabricated gauge for sensors manufactured at Penn State. Because the physical vapor deposition process is more precise than the etching process, the dimensionality of the platinum is more controlled than the copper (as identified by a slight overhang of platinum on the top portion of the gauge). Although these gauges represent a sensing element 90 µm × 500 µm, gauges with features as small as 4 µm have been successfully fabricated at Penn State. Controlling the geometry and the amount of platinum deposited changes the resistance of the sensor. After the fabrication is complete, the sensor must be calibrated before it can be used.

Fig. 2
Comparison of HFG design intent with as-built geometry
Fig. 2
Comparison of HFG design intent with as-built geometry
Close modal

Calibration

Traditionally, the calibration process has been performed using a calibration bath. This study evaluates a standard calibration bath process and expands the calibration procedures by proposing the feasibility of another technique called the 3-omega method, which enables continuous-duration facilities to routinely check HFG calibration coefficients.

Oil Bath Calibration.

Because platinum has a linear resistance behavior over the temperatures of interest, the calibration curve is dependent upon the offset and the slope of the fit. The offset is the reference resistance and the slope of the line is the coefficient of resistance. For this study, a temperature calibration bath was used with silicone oil as the heat transfer fluid. The bath has a uniformity to within 0.02 °C and a stability within 0.01 °C. A higher degree of accuracy was achieved by pairing the bath with a reference thermometer and an RTD, ultimately yielding an overall accuracy of 0.05 °C.

Before the calibration procedure begins, gauges were annealed at a fixed temperature, slightly higher than the expected maximum operating temperature of the sensor, until the nominal resistance of the gauges changed by less than 1% over an hour. This step is necessary to relax the internal stresses in the gauge left during the evaporative deposition process and mitigate hysteresis in thermal properties.

Next, the calibration bath was automatically swept through the predefined temperature calibration range and repeated several times to quantify hysteresis; typical results are shown in Fig. 3. The calibration measures the voltage drop across the lead wires as well as the platinum element. Calibrating using the same leads as in the experiment ensures lead effects are taken into account. The resistance is then calculated by dividing by a known current excitation. If each calibrated gauge had an R2 value of greater than 0.999, corresponding to a maximum temperature error of 0.3 °C. The calibration curves in Fig. 3 represent five different gauges made from three different institutions. The sensitivity of the gauge is directly dependent on the coefficient of resistance [10], so some users find it advantageous to maximize this quantity. In this case, different levels of deposited platinum directly contribute to the coefficient of resistance: when more platinum is deposited, the coefficient of resistance increases. For deposition thicknesses greater than approximately 100 nm, quantum effects are expected to become negligible [33]. However, Table 2 shows that the coefficient of resistance for these gauges is still far from the bulk value for platinum.

Fig. 3
Calibration curve for five different thin film RTDs. The slope of the line represents the coefficient of resistance.
Fig. 3
Calibration curve for five different thin film RTDs. The slope of the line represents the coefficient of resistance.
Close modal
Table 2

Coefficient of resistance for different sensors

ManufacturerPt depositionCoefficient of resistance (β) (°C−1)
Penn StateLow (50 nm)7.08E-4
Penn StateHigh (200 nm)1.84E-3
AFRLLow1.45E-3
AFRLHigh1.62E-3
Oxford1.67E-3
Platinum [34]3.85E-3
ManufacturerPt depositionCoefficient of resistance (β) (°C−1)
Penn StateLow (50 nm)7.08E-4
Penn StateHigh (200 nm)1.84E-3
AFRLLow1.45E-3
AFRLHigh1.62E-3
Oxford1.67E-3
Platinum [34]3.85E-3

There are two primary explanations for the discrepancies identified in Table 2 compared with the expected value for bulk platinum. First, the platinum may be improperly annealed, as temperatures higher than the polyimide decomposition point are required to achieve a fully annealed state [35]. Second, the polyimide surface roughness may contribute. For example, platinum thin-film RTDs on the order of 350 nm deposited on silicon showed higher coefficients of resistance closer to the bulk values [36]. Ultimately, although sensitivity increase may be desirable, gauges with coefficients of resistance similar to Table 2 have been used successfully for several decades.

3-Omega Calibration Check.

Because continuous-duration facilities are subject to long run times, the potential for gauge erosion over time necessitates an in situ calibration capability. If the gauge calibration shifts for any reason during a test, it would not be feasible to recalibrate in an oil bath before and after test campaigns due to the extended times required to remove and install the necessary hardware.

To address this need, the 3-omega technique [37] states that exciting a properly designed gauge with a sinusoidal current at frequency ω will introduce Joule heating and a corresponding temperature rise at a harmonic frequency, 2ω. In turn, this creates a voltage across the resistor that has a dominating 1ω component with a small 3ω component. By independently measuring the amplitudes of these harmonic components, Eq. (1) can be applied.
(1)
which can be solved numerically. This solution is presented in Fig. 4 with a plot of thermal wavelength given by
(2)
Fig. 4
Numerical solution to the third harmonic of voltage solved as a function of a given heating frequency
Fig. 4
Numerical solution to the third harmonic of voltage solved as a function of a given heating frequency
Close modal

To achieve accurate measurements of thermal conductivity with this method, the thermal wavelength should be less than the thickness of the material to satisfy the semi-infinite boundary condition requirement [38]. In this study, the thickness of the etched Pyralux (EP) was 50 µm. Based on these dimensions, the dashed vertical lines in Fig. 5 illustrate the region where the linear approximation is most appropriate for the current gauge design.

Fig. 5
Results using the 3-omega method to measure thermal conductivity
Fig. 5
Results using the 3-omega method to measure thermal conductivity
Close modal
Simplifying Eq. (1) for the linear region of the solution, Eq. (3) is obtained with a log-base slope dependent upon the thermal conductivity of the polyimide and the coefficient of resistance of the heater.
(3)
In the above equation, Vh,03 is measured, and Rh,0, L, and bh are known from the gauge characterization. The thermal properties ρ and cp are also considered known parameters (more details on these measurements are offered later in the paper). The frequency ω is controlled experimentally, therefore leaving two unknowns: βh and k. If one of these parameters is known, then the technique can be used to solve for the other.

The technique offers two primary benefits. First, because the coefficient of resistance is known through a traditional oil bath calibration, the 3-omega method can be used to obtain thermal conductivity across a range of temperatures. Second, with the knowledge of thermal conductivity as a function of temperature, the same method can be subsequently used to obtain the in situ coefficients of resistance, assuming the thermal properties of the substrate do not change over time. Work is currently being conducted to verify this assumption. This feature allows for the HFGs to be easily calibrated before, during, and after each measurement.

To test this theory, thermal conductivity was measured using the 3-omega method at room temperature using the gauge shown in Fig. 2. The 3-omega technique was applied over a range of frequencies, and the results are reported in Fig. 5. In Fig. 5, the dashed lines represent the analytical solution of both the in-phase and out-of-phase signals. To obtain these curves, the thermal conductivity of the material was assumed to be a fixed known value representative of EP. Using this value as a standard, the thermal conductivity determined from the 3-omega method differed by as much as ∼20%. This discrepancy is attributed to a narrow range of frequencies used to obtain the results, which will be improved by a gauge geometry change for future tests. By having a larger range of frequencies to test, a more representative thermal conductivity can be found.

Although this method is still in its preliminary stages of development for HFG use, it has been used in for decades in the nanofabrication field to characterize the thermal conductivity of thin films. Potential benefits exist by leveraging this established technique with HFG operation, especially in continuous-duration facilities.

Thermal Property Determination

The material properties required to transform measured temperature to heat flux for a double-sided type gauge are density, specific heat, and thermal conductivity, as well as the thickness of the dielectric. More precisely, the heat flux depends upon the thermal product (ρcpk) and the ratio k/d. Quantifying these parameters and understanding the associated uncertainties is necessary to bound the corresponding uncertainty in the heat flux calculations. The properties presented here represent measurements of EP, and comparisons are drawn with other polyimides from previous studies.

Specific heat measurements were obtained using differential scanning calorimetry following ASTM E1269 [39]. The overall uncertainty for this method was 10%. Figure 6 shows the specific heat as a function of temperature, which exhibits a 45% variation over the range of temperatures tested and aligns well with measurements from previous studies. In previous studies, Choy et al. [40] measured Kapton H while Kotel’nikov [41] and Lambert [42] measured generic polyimide film. In Fig. 6, the value for EP reported by the manufacturer effectively represents the mean value across the selected temperature range but does not appropriately account for variations with temperature. The density of the EP was measured following ASTM D6093 [43]. This volume displacement measurement found the density to be 1510 kg/m3 at 20 °C with an accuracy of 2%. Because the thermal coefficient of expansion for these polyimide films is known with high stability, it is possible to approximate changes of density as a function of temperature using the coefficients shown in Table 3.

Fig. 6
Specific heat as a function of temperature for EP compared with other studies
Fig. 6
Specific heat as a function of temperature for EP compared with other studies
Close modal
Table 3

Coefficient of thermal expansion based upon kapton HN

Thermal coefficient of expansion (ppm/°C)Temperature range (°C)
αT11730–100
αT232100–150
Thermal coefficient of expansion (ppm/°C)Temperature range (°C)
αT11730–100
αT232100–150
Because the mass of the sample remains unchanged, the density change can be correlated with volume using Eq. (4) assuming isotropic expansion.
(4)
Choosing ΔT1 = 70 °C and ΔT2 = 50 °C, the corresponding volume change is less than 1%, and serves as a first-order direct approximation of density variations over this temperature range, which represents a variation much less than the other properties of interest.

Thermal conductivity measurements were collected using ISO Standard 22007-2. Three measurements were collected: Kapton HN at 23 °C, EP at 23 °C, and EP at 150 °C, and these results are reported in Fig. 7. The absolute uncertainty of the measurement is 5%, and the relative standard deviation was less than 2.3%. Although only two points were measured for EP, the thermal conductivity of the samples varied by more than 27% over the temperature range of interest.

Fig. 7
Thermal conductivity measurements for EP as a function of temperature
Fig. 7
Thermal conductivity measurements for EP as a function of temperature
Close modal

After each thermal property was individually characterized, the thermal product (ρcpk) was examined. Table 4 compares the present study with measurements at two temperatures to previously published data from AFRL [27], Oxford [23,24], MIT [3], and VKI [32] as well as manufacturer-reported specifications [4547]. Although this table shows a wide range of values, this spread is not unexpected. Each institution characterized their own materials, and the methods for quantifying thermal properties varied between studies. In the studies from VKI and Oxford, the material the gauge was mounted upon was taken into account, which is why Table 4 has multiple studies or values from the same institution. Details about the methods can be obtained from the individual references. Although the dielectric was always a polyimide, manufacturers and batch-to-batch variations contribute noticeably to the result. Based on this observation, it is increasingly important for users to regularly measure the thermal properties of the dielectric for each sensor build to obtain high-quality heat flux measurements.

Table 4

Comparison of thermal properties from various studies used to calculate heat flux through thin polyimide films

InstitutionReferenceMaterialρ (kg/m3)cp(J/kgK)k (W/mK)ρcpk(J/Km2s0.5)% Diff relative to Tprop = 50 °C% Diff relative to Tprop = 150 °CUncertainty ρcpk
Present study (50 °C)Etched Pyralux15108760.1949630.85%5.89%
Present study (150 °C)Etched Pyralux151012430.2264930.85%6.25%
Air Force (2011) (23 °C)[27]Kapton HN141010580.185184.44%20.18%6.52%
Oxford (1999)[24]Upilex4950.00%23.73%4.20%
Oxford (1986)[23]Kapton HN56012.90%13.71%
MIT (1985)[3]Kapton HN57515.93%11.40%5.00%
VKI (2002)[32]Upilex699–73147.37%12.63%8.80%
DuPont[44]Etched Pyralux143010890.2663628.23%2.00%
DuPont[45]Kapton HN142010900.1243113.10%33.59%
UBE[46]Upilex-S147011300.2964930.85%0.00%
InstitutionReferenceMaterialρ (kg/m3)cp(J/kgK)k (W/mK)ρcpk(J/Km2s0.5)% Diff relative to Tprop = 50 °C% Diff relative to Tprop = 150 °CUncertainty ρcpk
Present study (50 °C)Etched Pyralux15108760.1949630.85%5.89%
Present study (150 °C)Etched Pyralux151012430.2264930.85%6.25%
Air Force (2011) (23 °C)[27]Kapton HN141010580.185184.44%20.18%6.52%
Oxford (1999)[24]Upilex4950.00%23.73%4.20%
Oxford (1986)[23]Kapton HN56012.90%13.71%
MIT (1985)[3]Kapton HN57515.93%11.40%5.00%
VKI (2002)[32]Upilex699–73147.37%12.63%8.80%
DuPont[44]Etched Pyralux143010890.2663628.23%2.00%
DuPont[45]Kapton HN142010900.1243113.10%33.59%
UBE[46]Upilex-S147011300.2964930.85%0.00%

Beyond the discrete results provided for the present study in Table 4, the combination of measured properties over a range of temperatures yields thermal product for all temperatures between 50 °C and 100 °C. In this case, the thermal conductivity was only measured at two temperatures, but a linear approximation between those two points is appropriate [40]. The results in Fig. 8 are presented as a variation normalized by their mean values across the range of measurements. From this graph, it is easy to see that the specific heat contributes the most to the change in thermal product followed by the thermal conductivity. The density variations are an order of magnitude less than the other two parameters.

Fig. 8
Variation of thermal properties as a function of temperature, normalized by mean value over the temperature range
Fig. 8
Variation of thermal properties as a function of temperature, normalized by mean value over the temperature range
Close modal

Based on the variations identified in Fig. 8, it is important to quantify thermal properties at the temperature of operation to reduce errors. This importance increases for continuous-duration facilities that may have a larger variance of local gauge operating temperatures. In Table 4, Oxford, MIT, and VKI implicitly obtained thermal properties as lumped parameters, but discussions of varying temperatures were not included. In contrast, individual parameter measurements in the present study provide an inherent ability to control the temperature.

Error Sensitivity Analysis

Although this study considers both the single- and double-sided gauges, double-sided gauges must be used in a continuous-duration facility because the backside (metal) temperature will not fulfill a semi-infinite assumption. For this reason, the subsequent error sensitivity analysis was completed for a double-sided gauge only.

The uncertainty analysis was completed using a perturbation method in conjunction with Oldfield’s impulse response theory (IRT) toolbox [48]. This toolbox takes the temperature signals from the gauge, Ttop and Tbot, as well as the material properties as inputs. The output is the corresponding heat flux. Since the thermal properties are a function of temperature, it is advantageous to quantify the uncertainty in the output, q, for thermal properties at different temperatures.

To this point, calculated heat flux has been introduced as a function of the top and bottom gauge temperatures, the thermal product, and the thermal conductivity over the thickness of the substrate. As shown in the last section, the thermal conductivity and specific heat are direct functions of temperature. To obtain accurate measurements, the temperature of the EP during test operation must be known, and the operating temperature must fall within the temperature range over which the thermal properties were characterized. To illustrate the importance of these steps, a synthetic test signal is considered whose data are captured at an EP temperature of 100 °C and creates a constant heat flux of 50,000 W/m2 as shown in Fig. 9. The temperature traces come from analytical solutions while the thermal properties come from experimental conditions. By using analytical temperature traces, the influence of thermal properties can be isolated from signal noise and an imperfect experimental setup.

Fig. 9
Analytical temperature traces from two sides of a double-sided gauge corresponding to a constant heat flux
Fig. 9
Analytical temperature traces from two sides of a double-sided gauge corresponding to a constant heat flux
Close modal
To understand the influence of thermal property variations with temperature on calculated heat flux, the same temperature traces are considered (i.e., with a substrate temperature of 100 °C), but the results were processed (with IRT) using thermal properties of the EP substrate quantified at temperatures different from the operating temperature. Following this process, Fig. 10 shows the percent error ɛ introduced by variations of thermal properties with temperature, defined by
(5)
The abscissa in Fig. 10 is denoted as TopTprop where Top is the operating temperature of the gauge and Tprop is the evaluation temperature of the thermal properties. Based on Fig. 10, the operating temperature must be within ± 10 °C of the temperature at which properties were measured to achieve a corresponding error less than 3%. As a corollary to this statement, if a discrepancy between operating temperature and characterization temperature of 50 °C occurs, an error of up to 14% may be introduced. This error arises because the temperature traces used to create the constant heat flux will alter the shape of the resulting heat flux when processed with different thermal properties. As the trend of a thermal product with temperature shown in Fig. 10 is approximately linear, the corresponding influence on error propagated through the processing algorithm yields a pattern in Fig. 10 that is approximately symmetric. The results in Fig. 10 were also evaluated for different levels of mean heat flux and different operating temperatures, yielding negligible differences on the magnitude or trend of error as a function of temperature.
Fig. 10
Error in the quantified heat flux based on a temperature difference between gauge operating temperature and thermal property evaluation temperature
Fig. 10
Error in the quantified heat flux based on a temperature difference between gauge operating temperature and thermal property evaluation temperature
Close modal

In this assessment, the error quantified in Fig. 10 is independent of the accuracy of the measured properties themselves. Using information from the previous sections, the uncertainties of thermal conductivity (k), specific heat (cp), and density (ρ) are 5%, 10%, and 2%, respectively. Other factors include uncertainty of measured temperature, dielectric thickness (believed to be 10% [47]), and calibration coefficient of resistance of the RTD (0.5%). Assuming that the thermal properties are known for the temperature at which data is recorded (i.e., independent of the errors introduced in Fig. 10), the overall effect on the calculation of heat flux from the above uncertainties is approximately 8% and was calculated using a perturbation method [49]. A combination of these relative errors would be applied for each individual application to calculate a representative overall uncertainty.

Steady Facility Post-Processing

Although many post-processing schemes for double-sided heat flux gauges exist, the IRT introduced by Oldfield [48] is widely used for its low numerical error and computational efficiency. However, this method assumes that the measurements start from a zero condition, dictating an inherent settling time that appears when processing data beginning from a steady-operating condition. This settling time is derived from the Laplace transform and one-sided z-transform both starting at t = 0 in the IRT analysis. Therefore, to apply this processing technique for continuous-operation facilities, the settling time and the associated errors must be quantified by comparing the analytical and numerical solutions. Although the error has been studied previously for a step-change in heat flux (i.e., short-duration blow-down tests), a step change does not correlate well to continuous-duration facility operation. To more accurately represent the conditions in a continuous-duration flow facility, synthetic temperature traces were created which correspond to a sinusoidal heat flux around a known mean value, and these signals were filtered using Oldfield’s toolbox for comparison with the analytical solution. Starting with the boundary condition for heat flux experienced by the gauge,
(6)
where x = 0 is the top (air side) of the gauge and x = d at the back (metal side), an analysis was performed for a sensor neglecting glue layers and protective layers on top of the gauge.

Synthetic temperature profiles for the top and bottom of the gauge were found solving the unsteady conduction equation. A full analysis of the analytical solution can be found in previous work [3]. Figure 11 shows the result comparing the analytical solution with processed data using the IRT method for a single sinusoidal frequency ϖ. In Fig. 11, the analytical solution is represented by the dashed line. For times at the beginning of the processed data window, there is a discrepancy between the heat flux calculated through Oldfield’s IRT method (solid line) and the analytical solution. Because the IRT solution asymptotically approaches the analytical solution, the time at which one full period is within 1% of the analytical solution is denoted tɛ1.

Fig. 11
Impulse response processing for a sinusoid
Fig. 11
Impulse response processing for a sinusoid
Close modal

Following the procedure outlined in Fig. 11, a series of cases were tested over a frequency range from 0 to 50 kHz (covering several harmonics of potential blade passing frequencies). The Q0 and Qmean values were varied as seen in Table 5. These were chosen to account for different fluctuation levels in the signal.

Table 5

Parameters tested for IRT under steady operation

ParameterTested values
ϖ(kHz)0–50
Qmean (W/m2)[2, 20, 2 × 105]
Q0Qmean(W/m2)[0.01, 0.1, 0.5]
ParameterTested values
ϖ(kHz)0–50
Qmean (W/m2)[2, 20, 2 × 105]
Q0Qmean(W/m2)[0.01, 0.1, 0.5]

For all test cases, the settling time to reach an error below 1% was less than 0.01 s over the entire range of frequencies. These results, shown in Fig. 12, are a function of the ratio of the amplitude to the mean value, with an inverse relationship between heat flux amplitude and calculated settling time. Phase lag and RMS error were also calculated with influences less than 0.5% for all values tested. Because all the error metrics tested were found to be less than 0.5% after the settling time, this method of data processing can be applied for continuous-duration facilities if the user collects more data than necessary and identifies that an initial portion must be removed to account for the settling time associated with the processing technique.

Fig. 12
Correction time for 1% error versus heating frequency
Fig. 12
Correction time for 1% error versus heating frequency
Close modal

Impinging Jet Experimental Setup

After characterizing the individual properties of the gauges, a shutter-rig-type experiment was setup to test the whole gauge at the device level. This test apparatus consists of a heat gun mounted at a fixed distance from an HFG sensor. The heat gun was experimentally characterized as a function of device control settings (outlet temperature and flow level) with the use of a pitot-static probe and thermocouple to measure the exit conditions of the device. Through this process, three Reynolds numbers based on nozzle diameter ReD were selected for assessment: 5 × 103, 10 × 103, and 12 × 103.

To quantify the heat flux from the heat gun, a Vatell HFM was mounted at a spacing of 4D from the outlet of the heat gun, as shown in Fig. 13. This sensor has an uncertainty of 5% and was used as the reference in this study [50], and the spacing was chosen to compare with the literature [51]. Results from the impinging jet at all ReD were found to be within 10% of values for ReD = 10 × 103 and ReD = 12 × 103 reported in the literature study.

Fig. 13
Diagram of the experimental setup for shutter-rig tests where (a) shows the reference gauge (Vatell setup) and (b) shows how the HFGs were mounted (HFG setup)
Fig. 13
Diagram of the experimental setup for shutter-rig tests where (a) shows the reference gauge (Vatell setup) and (b) shows how the HFGs were mounted (HFG setup)
Close modal

Comparison of Gauge Performance

Three different thin film heat flux gauges were used under the shutter rig. The first was a double-sided thin film heat flux gauge manufactured by AFRL. The second and third were a single-sided gauge manufactured at Penn State and Oxford. Note that because of the experimental setup, both double- and single-sided gauges could be used. Although double-sided gauges must be used in a steady facility, for a transient jet, both are applicable.

Each gauge was excited using a current amplitude selected to yield a voltage drop across the resistor of 0.25 V at room temperature. Because the single-sided gauges require accurate thermal properties of the material onto which the gauges are mounted, the gauges were adhered onto an aluminum block whose thermal properties were measured using the same techniques outlined above. The aluminum block was instrumented with thermocouples to quantify the backside temperature of the block, as well as the temperature near the surface to compare to the thin film RTDs. When accounting for the uncertainty in the second layer thermal properties, the uncertainty in the measured heat flux from a single-sided PSU gauge is 10%, calculated by using the upper and lower bound for the thermal properties to determine the heat flux and comparing the difference.

These results, shown in Fig. 14, were averaged over 0.3 s of steady data after the shutter was released (sampled at 5 kHz). Also in Fig. 14 are values from Goldstein et al. who measured the heat flux from an impinging jet over a similar range of ReD using a temperature-sensitive liquid crystal technique (LCT) [51]. For each gauge, the heat flux was calculated using the thermal properties provided by the respective institution and reported in Table 4. Because the thermal properties of the Penn State gauge were known for a range of temperatures, the mean temperature of the sensor over the 0.3 s averaging window was used to calculate the heat flux, which was not possible with the single values from other institutions. To calculate the thickness of the sample, a three-dimensional microscope was used with a resolution of 4 µm. Combining this information, error bars in the figure were calculated using the quoted uncertainty from each institution.

Fig. 14
Comparison of gauges under impinging jet using shutter-rig apparatus with literature values
Fig. 14
Comparison of gauges under impinging jet using shutter-rig apparatus with literature values
Close modal

Overall, the gauge in closest agreement with the reference was Penn State. Because all three gauges were calibrated and excited with the same system, the errors between the gauges must be from the thermal properties used to process the data. Since the Penn State property determination came from a piece of EP near where the gauge was built, it was potentially more representative of the dielectric than the values quoted by the other institutions. For this reason, this specific gauge read closer to the commercially available gauge. This statement reiterates the importance of quantifying thermal properties for each set of gauge builds.

Conclusions

This study addresses the development and implementation of thin-film heat flux gauges specific to a continuous-duration test facility, such as the Steady Thermal Aero Research Turbine Laboratory at Penn State. The fabrication of the gauges has been outlined and shown to have alignment within 5 µm which is critical to the orthogonal 1D assumption in a steady facility. Although the fabrication is an important step, gauge deterioration is a major concern for steady-operating facilities.

To address this concern, unique calibration procedures involving the 3-omega method show utility as a method to assess gauge deterioration and perform in situ calibration verifications. This method enables more frequent calibrations, thereby lowering uncertainty in the measurements associated with long run times.

Thermal property determination, which represents a significant contribution of error in heat flux measurements, was addressed by independently measuring the density, specific heat, and thermal conductivity over a range of temperatures. These measurements showed an overall thermal product variation of more than 30% over the range of temperatures studied. Using this information, an analytical solution to a step-change in the heat flux was used to determine the error in the heat flux measurement introduced by evaluating heat flux with thermal properties quantified at a temperature different from the sensor operating temperature through impulse response theory. Based on these measurements with polyimide films, thermal properties should be measured within 10 °C of desired test conditions to maintain a propagated error of less than 3%.

For post-processing, the IRT method was used for its computational efficiency. Post-processing errors from the IRT method were calculated using an analytical solution to a sinusoidal heat flux. Over a range of frequencies, the associate settling time to achieve a processed error less than 1% is less than 0.01 s, and the settling time is independent of mean heat flux level. The RMS and mean value errors for the method were also found to create less than a 0.5% error over all frequencies.

Finally, the mean heat flux from an impinging jet was measured using thin-film heat flux gauges from Penn State, AFRL, and Oxford. All gauges show agreement with a defined standard that was within their uncertainty over all conditions tested. Because the accuracy of these gauges depends on thermal property determination, the importance of correct thermal property determination is shown exemplifying that to obtain accurate measurements, thermal properties must be checked often for the specific polyimide beneath the gauge.

Beyond steady facilities, the process outlined in this paper shows the course of HFG implementation in any steady environment, including gas turbine engines themselves. Although the fabrication techniques must advance before such implementation, the process of calibration, thermal property determination, and data post-processing can stay the same to those outlined in this paper.

Acknowledgment

The authors thank the staff at Penn State’s Nanofabrication Lab. This work would not have been possible without their knowledge, specifically Michael Labella, Kathy Gehoski, and Ted Gehoski. Appreciation is also given to Professor Martin Oldfield whose Impulse Response Method was used to process the heat flux data and who gave insight to the inherent settling time that goes with the process. Finally, the authors are grateful to Carlos Vargas Venegas who pioneered the thin film thermal property techniques at Penn State and Brian Siroka, who ran many shutter-rig tests to obtain some of the data presented in this work.

Funding

The authors would like to recognize and thank Pratt & Whitney and the U.S. Department of Energy National Energy Technology Laboratory under Award Number DE-FE0025011 for supporting research presented in this paper. This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof.

Nomenclature

b=

resistive element half-width

d=

thickness

k=

thermal conductivity

q=

heat flux

t=

time

D=

diameter

H=

spacing from impingement jet, H = 4D

I=

current

L=

length of heater

Q=

sinusoidal heat flux

R=

resistance

T=

static temperature

V=

voltage

q′=

wave number = i/λ

cp=

specific heat

Pt=

platinum

ReD=

Reynolds number, ReD = U·D·ν−1

Greek

α=

linear coefficient of thermal expansion

β=

coefficient of resistance

Δ=

change in quantity

ɛ=

percent error

=

3-omega constant = 0.922

λ=

thermal wavelength

ρ=

density

ω=

excitation frequency for 3-omega method

ϖ=

frequency of boundary condition for the IRT method

Subscripts

bot=

bottom side of heat flux gauge

h=

heater

mean=

average value

op=

evaluated at operating condition

prop=

thermal property evaluation condition

ref=

reference temperature

top=

top side of heat flux gauge

0=

peak amplitude value

=

amplitude of the third harmonic of ω

References

1.
Jones
,
T. V.
, and
Schultz
,
D. L.
,
1973
, “Heat-Transfer Measurements in Short-Duration Hypersonic Facilities,”
AGARD-AG-165
,
Technical Editing and Reproduction Ltd.
,
London
.
2.
Chana
,
K.
,
Cardwell
,
D.
, and
Jones
,
T.
,
2013
, “
A Review of the Oxford Turbine Research Facility
,”
Proceedings of the ASME Turbo Expo 2013: Turbine Technical Conference and Exposition. Volume 3C: Heat Transfer
,
San Antonio, TX
,
June 3–7, 2013
, p.
V03CT14A026
.
3.
Epstein
,
A. H.
,
Guenette
,
G. R.
,
Norton
,
R. J. G.
, and
Yuzhang
,
C.
,
1986
, “
High-Frequency Response Heat-Flux Gauge
,”
Rev. Sci. Instrum.
,
57
(
4
), pp.
639
649
. 10.1063/1.1138882
4.
Mathison
,
R. M.
,
Wishart
,
M. B.
,
Haldeman
,
C. W.
, and
Dunn
,
M. G.
,
2011
, “
Temperature Predictions and Comparison With Measurements for the Blade Leading Edge and Platform of a 1 1/2 Stage Transonic HP Turbine
,”
ASME J. Turbomach.
,
134
(
1
), p.
011016
. 10.1115/1.4002992
5.
Iliopoulou
,
V.
,
Deńos
,
R.
,
Billiard
,
N.
, and
Arts
,
T.
,
2004
, “
Time-Averaged and Time-Resolved Heat Flux Measurements on a Turbine Stator Blade Using Two-Layered Thin-Film Gauges
,”
ASME J. Turbomach.
,
126
(
4
), pp.
570
577
. 10.1115/1.1791647
6.
Nasir
,
S.
,
Carullo
,
J. S.
,
Ng
,
W.
,
Thole
,
K. A.
,
Wu
,
H.
,
Zhang
,
L. J.
, and
Moon
,
H. K.
,
2009
, “
Effects of Large Scale High Freestream Turbulence and Exit Reynolds Number on Turbine Vane Heat Transfer in a Transonic Cascade
,”
ASME J. Turbomach.
,
131
(
2
), p.
021021
. 10.1115/1.2952381
7.
Clark
,
J. P.
,
Polanka
,
M. D.
,
Meininger
,
M.
, and
Praisner
,
T. J.
,
2003
, “
Validation of Heat-Flux Predictions on the Outer Air Seal of a Transonic Turbine Blade
,”
ASME J. Turbomach.
,
128
(
3
), pp.
589
595
.10.1115/1.2184351
8.
Anthony
,
R. J.
,
Jones
,
T. V.
, and
LaGraff
,
J. E.
,
2004
, “
High Frequency Surface Heat Flux Imaging of Bypass Transition
,”
ASME J. Turbomach.
,
127
(
2
), pp.
241
250
. 10.1115/1.1860379
9.
Dunn
,
M.
, and
Mathison
,
R.
,
2013
, “
History of Short-Duration Measurement Programs Related to Gas Turbine Heat Transfer, Aerodynamics, and Aeroperformance at Calspan and the Ohio State University
,”
ASME J. Turbomach.
,
136
(
4
), p.
041004
. 10.1115/1.4024898
10.
Epstein
,
A. H.
,
Guenette
,
G. R.
, and
Norton
,
R. J. G.
,
1985
, “
The Design of the MIT Blowdown Turbine Facility
,”
MIT Gas Turbine Lab Report 174
. https://dspace.mit.edu/bitstream/handle/1721.1/104724/16206191.pdf?sequence=1
11.
Sieverding
,
C. H.
, and
Arts
,
T.
,
1992
, “
The VKI Compression Tube Annular Cascade Facility CT3
,”
Proceedings of the ASME 1992 International Gas Turbine and Aeroengine Congress and Exposition. Volume 5: Manufacturing Materials and Metallurgy; Ceramics; Structures and Dynamics; Controls, Diagnostics and Instrumentation; Education
,
Cologne, Germany
,
June 1–4, 1992
, p.
V005T16A001
,
ASME
,
New York
.
12.
Dunn
,
M. G.
,
George
,
W. K.
,
Rae
,
W. J.
,
Woodward
,
S. H.
,
Moller
,
J. C.
, and
Seymour
,
P. J.
,
1986
, “
Heat-Flux Measurements for the Rotor of a Full-Stage Turbine: Part II—Description of Analysis Technique and Typical Time-Resolved Measurements
,”
ASME J. Turbomach.
,
108
(
1
), pp.
98
107
. 10.1115/1.3262030
13.
Guenette
,
G. R.
,
Epstein
,
A. H.
,
Giles
,
M. B.
,
Haimes
,
R.
, and
Norton
,
R. J. G.
,
1989
, “
Fully Scaled Transonic Turbine Rotor Heat Transfer Measurements
,”
ASME J. Turbomach.
,
111
(
1
), pp.
1
7
.10.1115/1.3262231
14.
Haldeman
,
C. W.
, and
Dunn
,
M. G.
,
2004
, “
Heat-Transfer Measurements and Predictions for the Vane and Blade of a Rotating High-Pressure Turbine Stage
,”
ASME J. Turbomach.
,
126
(
1
), pp.
101
109
.10.1115/1.1626132
15.
Kahveci
,
H. S.
,
Haldeman
,
C. W.
,
Mathison
,
R. M.
, and
Dunn
,
M. G.
,
2013
, “
Heat Transfer for the Film-Cooled Vane of a 1-1/2 Stage High-Pressure Transonic Turbine—Part II: Effect of Cooling Variation on the Vane Airfoil and Inner End Wall
,”
ASME J. Turbomach.
,
135
(
2
), p.
021028
.10.1115/1.4006776
16.
Guo
,
S. M.
,
Spencer
,
M. C.
,
Lock
,
G. D.
,
Jones
,
T. V.
, and
Harvey
,
N. W.
,
1995
, “
The Application of Thin Film Gauges on Flexible Plastic Substrates to the Gas Turbine Situation
,”
Proceedings of the ASME 1995 International Gas Turbine and Aeroengine Congress and Exposition. Volume 4: Heat Transfer; Electric Power; Industrial and Cogeneration
,
Houston, TX
,
June 5–8, 1995
, p.
V004T09A049
,
ASME
,
New York
.
17.
Haldeman
,
C. W.
,
Dunn
,
M. G.
,
Barter
,
J. W.
,
Green
,
B. R.
, and
Bergholz
,
R. F.
,
2005
, “
Aerodynamic and Heat-Flux Measurements with Predictions on a Modern One and One-Half State High Pressure Transonic Turbine
,”
ASME J. Turbomach.
,
127
(
3
), pp.
522
531
. 10.1115/1.1861916
18.
Polanka
,
M. D.
,
Hoying
,
D. A.
,
Meininger
,
M.
, and
MacArthur
,
C. D.
,
2003
, “
Turbine Tip and Shroud Heat Transfer and Loading—Part A: Parameter Effects Including Reynolds Number, Pressure Ratio, and Gas-to-Metal Temperature Ratio
,”
ASME J. Turbomach.
,
125
(
1
), pp.
97
106
. 10.1115/1.1515338
19.
Polanka
,
M. D.
,
Clark
,
J. P.
,
White
,
A. L.
,
Meininger
,
M.
, and
Praisner
,
T. J.
,
2003
, “
Turbine Tip and Shroud Heat Transfer and Loading: Part B—Comparisons Between Prediction and Experiment Including Unsteady Effects
,”
Proceedings of the ASME Turbo Expo 2003, Collocated with the 2003 International Joint Power Generation Conference. Volume 5: Turbo Expo 2003, Parts A and B
,
Atlanta, GA
,
June 16–19, 2003
, pp.
691
702
,
ASME
,
New York
.
20.
Chana
,
K. S.
, and
Jones
,
T. V.
,
2003
, “
An Investigation on Turbine Tip and Shroud Heat Transfer
,”
ASME J. Turbomach.
,
125
(
3
), pp.
513
520
.10.1115/1.1575253
21.
Hager
,
J. M.
,
Simmons
,
S.
,
Smith
,
D.
,
Onishi
,
S.
,
Langley
,
L. W.
, and
Diller
,
T. E.
,
1991
, “
Experimental Performance of a Heat Flux Microsensor
,”
ASME J. Eng. Gas Turbines Power
,
113
(
2
), pp.
246
250
. 10.1115/1.2906555
22.
Ewing
,
J.
,
Gifford
,
A.
,
Hubble
,
D.
,
Vlachos
,
P.
,
Wicks
,
A.
, and
Diller
,
T.
,
2010
, “
A Direct-Measurement Thin-Film Heat Flux Sensor Array
,”
Meas. Sci. Technol.
,
21
(
10
), p.
105201
.10.1088/0957-0233/21/10/105201
23.
Doorly
,
J. E.
, and
Oldfield
,
M. L. G.
,
1986
, “
New Heat Transfer Gages for Use on Multilayered Substrates
,”
ASME J. Turbomach.
,
108
(
1
), pp.
153
160
.10.1115/1.3262015
24.
Piccini
,
E.
,
Guo
,
S. M.
, and
Jones
,
T. V.
,
2000
, “
The Development of a New Direct-Heat-Flux Gauge for Heat-Transfer Facilities
,”
Meas. Sci. Technol.
,
11
(
4
), pp.
342
349
. 10.1088/0957-0233/11/4/302
25.
Duan
,
F. L.
,
Li
,
J.
,
Gao
,
J.
,
Ding
,
G.
, and
Cao
,
X.
,
2018
, “
Integrated Fabrication of High-Temperature Microelectromechanical System Sensor on Aeroengine Turbine Blade
,”
J. Thermophys Heat Transfer
,
32
(
1
), pp.
1
4
. 10.2514/1.T5411
26.
Hodak
,
M.
,
2010
, “
Quantification of Fourth Generation Kapton Heat Flux Gauge Calibration Performance
,”
Electronic thesis
,
The Ohio State University
, https://etd.ohiolink.edu/pg_10?::NO:10:P10_ETD_SUBID:72618
27.
Anthony
,
R. J.
,
Clark
,
J. P.
,
Kennedy
,
S. W.
,
Finnegan
,
J. M.
,
Johnson
,
D.
,
Hendershot
,
J.
, and
Downs
,
J.
,
2011
, “
Flexible Non-Intrusive Heat Flux Instrumentation for the AFRL Research Turbine
,”
Proceedings of the ASME 2011 Turbo Expo: Turbine Technical Conference and Exposition. Volume 5: Heat Transfer, Parts A and B
,
Vancouver, British Columbia, Canada
,
June 6–10
, pp.
1793
-
1803
,
ASME
,
New York
.
28.
Collins
,
M.
,
Chana
,
K.
, and
Povey
,
T.
,
2015
, “
New Technique for the Fabrication of Miniature Thin Film Heat Flux Gauges
,”
Meas. Sci. Technol.
,
26
(
2
), p.
025303
. 10.1088/0957-0233/26/2/025303
29.
Doorly
,
J. E.
, and
Oldfield
,
M. L. G.
,
1987
, “
The Theory of Advanced Multi-Layer Thin Film Heat Transfer Gauges
,”
Int. J. Heat Mass Transfer
,
30
(
6
), pp.
1159
1168
. 10.1016/0017-9310(87)90045-7
30.
De Cuyper
,
T.
,
Fossaert
,
G.
,
Collet
,
O.
,
Broekaert
,
S.
,
Chana
,
K.
,
De Paepe
,
M.
, and
Verhelst
,
S.
,
2015
, “
Calibration of a TFG Sensor for Heat Flux Measurements in a S.I. Engine
,”
SAE Int. J. Engines
,
8
(
4
), pp.
1692
1700
. 10.4271/2015-01-1645
31.
Zilles
,
D.
,
1999
, “
The Effect of Non-Isothermal Wall Boundary Temperature on Convective Heat Flux Over a Flat Plate
,”
thesis
,
The Ohio State University
.
32.
Billiard
,
N.
,
Iliopoulou
,
V.
,
Ferrara
,
F.
, and
Dénos
,
R.
,
2002
, “
Data Reduction and Thermal Product Determination for Single and Multi-Layered Substrate Thin-Film Gauges
,”
16th Symposium on Measuring Techniques in Transonic Supersonic Flow Cascades Turbomachines
,
Cambridge, UK
.
33.
Lacy
,
F.
,
2011
, “
Developing a Theoretical Relationship Between Electrical Resistivity, Temperature, and Film Thickness for Conductors
,”
Nanoscale Res. Lett.
,
6
(
1
), pp.
1
26
. 10.1186/1556-276X-6-636
34.
WIKA
,
2014
, “
Callendar-Van Dusen Equations for the Calibration of Platinum Resistance Thermometers
,”
WIKA data sheet IN 00.29
.
35.
Zhang
,
J.
,
Nagao
,
Y.
,
Kuwano
,
S.
, and
Ito
,
Y.
,
1997
, “
Microstructure and Temperature Coefficient of Resistance of Platinum Films
,”
Jpn. J. Appl. Phys.
,
36
(
2
), pp.
834
839
. 10.1143/JJAP.36.834
36.
Kreider
,
K. G.
,
Ripple
,
D. C.
, and
Kimes
,
W. A.
,
2009
, “
Thin-Film Resistance Thermometers on Silicon Wafers
,”
Meas. Sci. Technol.
,
20
(
4
), pp.
1
18
. 10.1088/0957-0233/20/4/045206
37.
Cahill
,
D. G.
,
1990
, “
Thermal Conductivity Measurement From 30 to 750 K: The 3ω Method
,”
Rev. Sci. Instrum.
,
61
(
2
), pp.
802
808
.10.1063/1.1141498
38.
de Koninck
,
D.
,
2008
, “
Thermal Conductivity Measurements Using the 3-Omega Technique: Application to Power Harvesting Microsystems
,”
thesis
,
McGill University
.
39.
American Society for Testing and Materials (ASTM®)
,
2011
, “
ASTM E1269—Standard Test Method for Determining Specific Heat Capacity by Differential Scanning
,”
West Conshohocken, PA
;
ASTM International
,
2018
.
40.
Choy
,
C. L.
,
Leung
,
W. P.
, and
Ng
,
Y. K.
,
1987
, “
Thermal Diffusivity of Polymer Films by the Flash Radiometry Method
,”
J. Polym. Sci. Part B Polym. Phys.
,
25
(
9
), pp.
1779
1799
.10.1002/polb.1987.090250901
41.
Kotel’nikov
,
G. V.
, and
Sidorovich
,
A. V.
,
1983
, “
Microcalorimeter Type DSM-2M for Use in the Study of Polymers
,”
Polym. Sci. U.S.S.R.
,
25
(
12
), pp.
3053
3059
. 10.1016/0032-3950(83)90060-6
42.
Brunco
,
D. P.
,
Thompson
,
M. O.
,
Otis
,
C. E.
, and
Goodwin
,
P. M.
,
1992
, “
Temperature Measurements of Polyimide During KrF Excimer Laser Ablation
,”
J. Appl. Phys.
,
72
(
9
), pp.
4344
4350
.10.1063/1.352198
43.
ASTM
,
2016
, “
ASTM-D6093 Standard Test Method for Percent Volume Nonvolatile Matter in Clear or Pigmented Coatings Using a Helium Gas Pycnometer
,”
West Conshohocken, PA
;
ASTM International
,
2016
.
44.
DuPont
,
2018
,
personal communication
.
45.
DuPont
,
2011
, “
DuPont Kapton HN Technical Data Sheet 5213
,” pp.
1
4
.
48.
Oldfield
,
M. L. G.
,
2008
, “
Impulse Response Processing of Transient Heat Transfer Gauge Signals
,”
ASME J. Turbomach.
,
130
(
2
), p.
021023
.10.1115/1.2752188
49.
Moffat
,
R. J.
,
1982
, “
Contributions to the Theory of Single-Sample Uncertainty Analysis
,”
ASME J. Fluids Eng.
,
104
(
2
), pp.
250
258
.10.1115/1.3241818
50.
Vatell Corporation
,
2015
, HFM-6, HFM-7, & HFM-8.
51.
Goldstein
,
R. J.
, and
Franchett
,
M. E.
,
1988
, “
Heat Transfer From a Flat Surface to an Oblique Impinging Jet
,”
ASME J. Heat Transfer
,
110
(
1
), pp.
84
90
.10.1115/1.3250477