In the experimental determination of the response time of a thermoresistive sensor to a radiation step, one needs to monitor the variation of the ohmic resistance of the sensor with time. This requires passing electrical current through the sensor, which introduces error in the measured response time due to the additional self-heating of the sensor by the electrical current. A theoretical method to eliminate this error is formulated and experimental results are presented. The proposed method also permits experimental determination of the heat capacity (mc) and heat transfer coefficient (UA). The sensitivity analysis of τ = mc/UA permits the optimal choice for the resistance measurement currents.

1.
Doebelin, E. O., 1976, Measurement System Applications and Design, McGraw-Hill, New York.
2.
Freire, R. C. S., Neto, J. S. R., Deep, G. S., Lima, A. M. N., and Lobo, P. C., 1994, Experimental Determination of the Time Constant of a Thermoresistive Sensor, IMTC’94 Conference Record Hamamatsu, Japan, May, pp. 450–453.
3.
Nelder
J. A.
, and
Mead
R.
,
1966
, “
A Simplex Method for Function Minimization
,”
The Computer Journal
, Vol.
7
, pp.
308
313
.
4.
Neto, J. S. R., 1993, “Ana´lise e Projeto de um Instrumento para Medic¸a˜o de Radic¸a˜ao,” Ph.D. thesis proposal, COPELE-UFPB, Campina Grande, PB, Brazil, Apr.
5.
Press, W. H., Flannery, B. P., Teuloksky, S. A., and Vetterling, W. T., 1988, Numerical Recipes in C—The Art of Scientific Computing, Cambridge University Press, Cambridge, UK.
This content is only available via PDF.
You do not currently have access to this content.