Magnetic bearings are becoming increasingly popular in ventricular assist devices. In most cases, blood fills a portion of the gap between the magnetic actuators and rotor. Understanding the effects of the operating fluid on magnetic suspension is necessary, particularly when the device geometry features a relatively large gap (typically defined as greater than um and as large as 2 mm in some pumps). A large gap reduces cell damage and allows unrestricted flow, it but increases the distance and amount of material through which magnetic fields must pass. These net effects of the operating fluid on the magnetic suspension can be characterized in terms of a contribution to the overall damping and stiffness of the system. This contribution may be caused by traditional tribological properties (density and viscosity) as well as the magnetic properties of the medium. This research isolates the effects of fluid diamagnetism on a magnetically levitated blood pump. Experimental transient and frequency responses of the system operating at 3000 to 6000 rpm are presented while pumping different liquid media: Blood, water, and fluids with similar densities and viscosities to each, but different magnetic susceptibilities. In order to calculate the stiffness and damping coefficients, a mathematical model of the system is iteratively updated to match experimental transient response. The resulting coefficients are validated in the frequency domain by means of simulations, which agree with experimental frequency response data. The relationship between diamagnetism, damping and stiffness coefficients of the test fluids with known magnetic susceptibility is also used to obtain a quantitative estimate of the magnetic susceptibility of blood.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
Diamagnetic Effects of Blood in a Magnetically Levitated Blood Pump
A. D. Gomez,
A. D. Gomez
Rochester Institute of Technology
, New York, USA
Search for other works by this author on:
S. W. Day
S. W. Day
Rochester Institute of Technology
, New York, USA
Search for other works by this author on:
A. D. Gomez
Rochester Institute of Technology
, New York, USA
S. W. Day
Rochester Institute of Technology
, New York, USAJ. Med. Devices. Jun 2009, 3(2): 027504 (1 pages)
Published Online: June 30, 2009
Article history
Published:
June 30, 2009
Citation
Gomez, A. D., and Day, S. W. (June 30, 2009). "Diamagnetic Effects of Blood in a Magnetically Levitated Blood Pump." ASME. J. Med. Devices. June 2009; 3(2): 027504. https://doi.org/10.1115/1.3134838
Download citation file:
663
Views
Get Email Alerts
Cited By
A Better Way to Visualize the Cervix: A Low-Cost Cervical Cancer Screening Device for LMIC Settings
J. Med. Devices (April 2025)
Comparative Survey-Based Study of Noninvasive Saliva Collection Devices
J. Med. Devices (April 2025)
Related Articles
Feasibility Study of MEMs Technique for Characterizing Magnetic Susceptibility of Subcellular Organelles
J. Med. Devices (June,2011)
Material Orientation Artifact Studies in Magnetic Resonance Imaging
J. Med. Devices (June,2009)
Elastic Stress and Magnetic Field Concentration Near the Vertex of a
Soft-Ferromagnetic 2D Compound Wedge
J. Appl. Mech (July,2008)
Study of MRI Susceptibility Artifacts for Nanomedical Applications
J. Nanotechnol. Eng. Med (November,2010)
Related Proceedings Papers
Related Chapters
Specific Device Examples
Magnetic Bearings for Mechanical Cardiac Assist Devices
Structure of Magnetic Actuator
Magnetic Bearings for Mechanical Cardiac Assist Devices
Modal Analysis of a New Magnetic Levitation Stage Based on Finite Element Simulation
International Conference on Mechanical Engineering and Technology (ICMET-London 2011)