The multiphotodiode array (MPA) is a novel transmission photoplethysmography (PPG) sensor to measure pulse wave velocity (PWV) in the finger. To validate the MPA, a setup was built to generate a red laser dot traveling over the MPA with known and constant scanning velocities. These scanning velocities were chosen to include speeds at least twice as high as those found in the normal range of PWV in healthy populations and were set at 12.9, 25.8, 36, or 46.7 m/s. The aim of this study was to verify the functionality of the MPA: performing local noninvasive PWV measurements. To illustrate the applicability of the MPA in clinical practice, an in vivo pilot study was conducted using the flow-mediated dilation (FMD) technique. The in vitro accuracy of the MPA was ±0.2%, 0.3%, 0.5%, and 0.6% at the applied scanning velocities. The MPA can measure PWVs with a maximum deviation of 3.0%. The in vivo pilot study showed a PWV before the FMD of 1.1±0.2 m/s. These results suggest that this novel MPA can reliably and accurately measure PWV within clinically relevant ranges and even well beyond.
Skip Nav Destination
Article navigation
March 2018
Research-Article
Design and Functional Testing of a Novel Blood Pulse Wave Velocity Sensor
Marit H. N. van Velzen,
Marit H. N. van Velzen
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: m.h.n.vanvelzen@tudelft.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: m.h.n.vanvelzen@tudelft.nl
Search for other works by this author on:
Arjo J. Loeve,
Arjo J. Loeve
Department of Biomechanical Engineering,
Faculty 3mE,
Delft University of Technology,
Mekelweg 2,
Delft 2628 CD, The Netherlands
e-mail: a.j.loeve@tudelft.nl
Faculty 3mE,
Delft University of Technology,
Mekelweg 2,
Delft 2628 CD, The Netherlands
e-mail: a.j.loeve@tudelft.nl
Search for other works by this author on:
Egbert G. Mik,
Egbert G. Mik
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: e.mik@erasmusmc.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: e.mik@erasmusmc.nl
Search for other works by this author on:
Sjoerd P. Niehof
Sjoerd P. Niehof
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: s.niehof@erasmusmc.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: s.niehof@erasmusmc.nl
Search for other works by this author on:
Marit H. N. van Velzen
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: m.h.n.vanvelzen@tudelft.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: m.h.n.vanvelzen@tudelft.nl
Arjo J. Loeve
Department of Biomechanical Engineering,
Faculty 3mE,
Delft University of Technology,
Mekelweg 2,
Delft 2628 CD, The Netherlands
e-mail: a.j.loeve@tudelft.nl
Faculty 3mE,
Delft University of Technology,
Mekelweg 2,
Delft 2628 CD, The Netherlands
e-mail: a.j.loeve@tudelft.nl
Egbert G. Mik
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: e.mik@erasmusmc.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: e.mik@erasmusmc.nl
Sjoerd P. Niehof
Department of Anesthesiology,
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: s.niehof@erasmusmc.nl
Laboratory of Experimental Anesthesiology,
Erasmus University Medical Center Rotterdam,
's-Gravendijkwal 230,
Rotterdam 3015 CE, The Netherlands
e-mail: s.niehof@erasmusmc.nl
1Corresponding author.
2Marit H. N. van Velzen and Arjo J. Loeve contributed equally to this work.
Manuscript received July 13, 2017; final manuscript received September 29, 2017; published online November 22, 2017. Assoc. Editor: Rafael V. Davalos.
J. Med. Devices. Mar 2018, 12(1): 011006 (7 pages)
Published Online: November 22, 2017
Article history
Received:
July 13, 2017
Revised:
September 29, 2017
Citation
van Velzen, M. H. N., Loeve, A. J., Mik, E. G., and Niehof, S. P. (November 22, 2017). "Design and Functional Testing of a Novel Blood Pulse Wave Velocity Sensor." ASME. J. Med. Devices. March 2018; 12(1): 011006. https://doi.org/10.1115/1.4038308
Download citation file:
Get Email Alerts
Cited By
Smart Kidney Stone Basket: A Proof of Concept With Digital Feedback Force Monitoring
J. Med. Devices (June 2025)
A Portable Ultrasound-Guided Surgical System for Breast and Thyroid Surgery
J. Med. Devices (June 2025)
The History, Innovation, and Standards of Intravenous Access Ports and Devices
J. Med. Devices (June 2025)
Related Articles
Data Communication Pathway for Sensing Guidewire at Proximal Side: A Review
J. Med. Devices (June,2017)
The Role of Computational Modeling and Simulation in the Total Product Life Cycle of Peripheral Vascular Devices
J. Med. Devices (June,2017)
Automated Real-Time Processing of Single Lead Electrocardiogram for Simultaneous Heart Rate and Respiratory Rate Monitoring
J. Med. Devices (June,2017)
Thrombogenicity Testing of Medical Devices in a Minimally Heparinized Ovine Blood Loop
J. Med. Devices (June,2017)
Related Proceedings Papers
Related Chapters
Generating Synthetic Electrocardiogram Signals Withcontrolled Temporal and Spectral Characteristics
Intelligent Engineering Systems through Artificial Neural Networks Volume 18
Fabrication of Silicon Carbide Photodiodes with Enhanced High-Temperature Performance by Electrochemical Anodization Technology
International Conference on Computer and Electrical Engineering 4th (ICCEE 2011)
Concluding remarks
Mechanical Blood Trauma in Circulatory-Assist Devices