A new compact centrifugal blood pump driven by a miniature DC servomotor has been designed for use for short-term extra corporeal and cardiac-assisted circulation. The impeller of the pump was connected directly to the motor by using a simple-gear coupling. The shaft for the impeller was sealed from blood by both a V-ring and a seal bearing. Either pulsatile or nonpusatile flow was produced by controlling the current supply to the motor. The pump characteristics and the degree of hemolysis were evaluated with regard to the configuration of the impeller with a 38-mm outer diameter in vitro tests; the impeller having the blade angles at the inlet of 20 deg and at the outlet of 50 deg was the most appropriate as a blood pump. The performance in an operation, hemolysis and thrombus formation in the pump were assessed by a left ventricular bypass experiment in dogs. It was suggested by this study that this prototype pump appears promising for use not only in animal experiments but also in clinical application.
Skip Nav Destination
Article navigation
August 1987
Research Papers
A Compact Centrifugal Blood Pump for Extracorporeal Circulation: Design and Performance
Shinobu Tanaka,
Shinobu Tanaka
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Search for other works by this author on:
Shuzo Yamamoto,
Shuzo Yamamoto
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Search for other works by this author on:
Ken-ichi Yamakoshi,
Ken-ichi Yamakoshi
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Search for other works by this author on:
Akira Kamiya
Akira Kamiya
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Search for other works by this author on:
Shinobu Tanaka
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Shuzo Yamamoto
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Ken-ichi Yamakoshi
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
Akira Kamiya
The Research Institute of Applied Electricity, Hokkaido University, Sapporo 060, Japan
J Biomech Eng. Aug 1987, 109(3): 272-278 (7 pages)
Published Online: August 1, 1987
Article history
Received:
March 31, 1986
Revised:
March 31, 1987
Online:
June 12, 2009
Citation
Tanaka, S., Yamamoto, S., Yamakoshi, K., and Kamiya, A. (August 1, 1987). "A Compact Centrifugal Blood Pump for Extracorporeal Circulation: Design and Performance." ASME. J Biomech Eng. August 1987; 109(3): 272–278. https://doi.org/10.1115/1.3138680
Download citation file:
Get Email Alerts
Related Articles
Analysis and Validation of a Unified Slip Factor Model for Impellers at Design and Off-Design Conditions
J. Turbomach (October,2011)
Optimization of an Innovative Rotary Shaft Pump (RSP)
J. Fluids Eng (November,2006)
Decomposition of Deterministic Unsteadiness in a Centrifugal Turbomachine: Nonlinear Interactions Between the Impeller Flow and Volute for a Double Suction Pump
J. Fluids Eng (January,2011)
Parametric Design of a Waterjet Pump by Means of Inverse Design, CFD Calculations and Experimental Analyses
J. Fluids Eng (March,2010)
Related Proceedings Papers
Related Chapters
Introduction
Design of Mechanical Bearings in Cardiac Assist Devices
Introduction I: Role of Engineering Science
Fundamentals of heat Engines: Reciprocating and Gas Turbine Internal Combustion Engines
Introduction
Mechanical Blood Trauma in Circulatory-Assist Devices